Characterization of MITLL CCID20-W93c2 thin 2k x 4k pixel size 15um x 15um new Laser annealing new AR coating tested by Lloyd Robinson Reported by M.Wei Aug.18,1998 "R"(A) amp worked only one very hot column TEST CONDITIONS temperature: -127c Voltages: Vdd = +19v Vr = +12.5v Vopg = 0v Vsub = 0v Vp1,2,3 = +2, -6v Vs1,2,3 = +6, -3v Vsw = +5, -5v RG = 10, 0v SCP = 10v read 3800(r)x2048(c) area only. changed gain of preamp. from 21 to 4.83 TEST RESULTS e-/DN: R: 0.42 Determined by Fe55 x-ray test serial CTE R: 0.999987 Determined by Fe55 x-ray test vertical CTE good (1.0) Determined by Fe55 x-ray test Vertical charge traps: 3 (in 3800(r) x 2048(c) area) Determined by charge-pumping method Read noise: R: 1.8e- DSC timing: 16us on data and 16us on baseline 2.8e- DSC timing: 4us on data and 4us on baseline Full well(nonmpp) 97 ke- Dark current: (@-127c) 1.51e-/pixel/hour QE(%)(average of big area) w.l. -127c 3200 8.0 3500 9.8 4000 37.9 4500 47.8 5000 58.2 6000 70.9 6500 73.6 7000 75.7 8000 60.4 9000 34.3 10000 5.6 Fringing 8000A 6.7% 9000A 19.1% 10000A 21.3% --------------------------------------------- QE variations of CCID20 w93c2 at -127c wavelength QE variation(%) 3200 47.8 3500 46.7 4000 6.4 4500 2.8 5000 0.3