Characterization of MITLL CCID20-W90c2 thin 2k x 4k pixel size 15um x 15um hz, new coating and annealing Reported by M.Wei Jul.9,1998 Both "R"(A) and "L"(B) amps worked TEST CONDITIONS temperature: -130c Voltages: Vdd = +19v Vr = +12.5v Vopg = 0v Vsub = 0v Vp1,2,3 = +2, -6v Vs1,2,3 = +6, -3v Vsw = +5, -5v RG = 10, 0v SCP = 10v read 3800(r)x2048(c) area only. changed gain of preamp. from 21 to 4.83 TEST RESULTS e-/DN: R: 0.40 Determined by Fe55 x-ray test L: 0.39 serial CTE R: 0.99994157 Determined by Fe55 x-ray test L: 0.99994245 Determined by Fe55 x-ray test vertical CTE 0.99999522 Determined by cosmic ray Vertical charge traps: 5 (in 2048(r) x 2048(c) area) Determined by charge-pumping method Read noise: R: 1.92e- DSC timing: 16us on data and 16us on baseline 3.04e- DSC timing: 4us on data and 4us on baseline L: 2.22e- DSC timing: 16us on data and 16us on baseline 2.77e- DSC timing: 4us on data and 4us on baseline Full well(nonmpp) 118 ke- Dark current: Measured from 1000s integration (nonmpp) 1.03e-/hour/pixel (@-110c) QE(%) -130c 3200A 16.4 3500A 22.2 4000A 44.5 4500A 54.4 5000A 65.5 6000A 77.7 6500A 82.7 7000A 81.4 8000A 67.7 9000A 35.8 10000A 6.1 Fringing 8000A 7.1% 9000A 5.5% 10000A 20.5% ---------------------------------------------