Characterization of MITLL CCID20-W12c1 thin 2k x 4k pixel size 15um x 15um low resistivity, old coating and laser annealing Reported by M.Wei Jul.9,1998 Both "R"(A) and "L"(B) amps worked Super hot column TEST CONDITIONS temperature: -130c Voltages: Vdd = +19v Vr = +12.5v Vopg = 0v Vsub = 0v Vp1,2,3 = +2, -6v Vs1,2,3 = +6, -3v Vsw = +5, -5v RG = 10, 0v SCP = 10v read 3800(r)x2048(c) area only. changed gain of preamp. from 21 to 4.83 TEST RESULTS e-/DN: R: 0.32 Determined by Fe55 x-ray test L: 0.34 serial CTE R: good Determined by Fe55 x-ray test L: good Determined by Fe55 x-ray test vertical CTE good Determined by cosmic ray Vertical charge traps: 5 (in 2048(r) x 2048(c) area) Determined by charge-pumping method Read noise: R & L noisy because of hot columns Full well(nonmpp) 101 ke- Dark current: QE(%) w.l. non treatment UV-in-O2 3200 2.6 2.3 3500 1.8 1.6 4000 29.8 30.5 4500 45.0 46.0 5000 60.7 68.4 6000 78.0 85.5 6500 81.4 85.7 7000 82.7 85.1 8000 62.6 63.2 9000 28.6 28.9 10000 6.0 5.9 Fringing 8000A 12.1% 9000A 6.9% 10000A 16.7% --------------------------------------------- QE variation with and without UV-flood (%) at -130c Without UV with UV 4000A 67.4 32.6 4500A 66.1 33.3 5000A 62.1 30.6 6000A 47.3 23.5 6500A 35.7 18.4 7000A 38.5 23.0 8000A 26.4 13.5 9000A 24.5 7.5 10000A 22.9 12.7