MIT/Lincoln Labs CCID20 W17C2 CCD Test Results

This phase 2 MIT/Lincoln Labs 2Kx4K CCD has the improved backside treatment, giving it a low amplitude brickwall at most wavelengths. A summary report is available. Noteworthy points about this CCD include:

Original postscript files are available from our anonymous ftp server and these provide better resolution and clarity than is usually possible on a web page. Here are a few figures to illustrate device highlights:


  • This graph shows the QE variations caused by the brick wall pattern. QE was measured at the peak and in the valley of the brick wall as well as over a large area to obtain an average QE. Over most wavelengths the variations are very small.Graph of quantum efficiency

  • Serial CTE as measured with Fe55 is show here. Readout was through the left or B amplifier. CTE was about 0.999990 per pixel. Parallel CTE was even better, at about 0.999965 per pixel.

    Plot showing Fe55 serial CTE measurement


    A defect near the serial register (at the top of this image) results in several blocked columns.