This phase 2 MIT/Lincoln Labs 2Kx4K CCD has the improved backside treatment, giving it a low amplitude brickwall at most wavelengths. A summary report is available. Noteworthy points about this CCD include:
Original postscript files are available from our anonymous ftp server and these provide better resolution and clarity than is usually possible on a web page. Here are a few figures to illustrate device highlights:

Serial CTE as measured with Fe55 is show here. Readout was through the left or B amplifier. CTE was about 0.999990 per pixel. Parallel CTE was even better, at about 0.999965 per pixel.

A defect near the serial register (at the top of this image) results in several blocked columns.