Lab Vacuum Equipment
The lab's high vacuum pumping station utilizes a turbomolecular dry pump
and a liquid nitrogen cold trap. Review the liquid
nitrogen SOP for proper handling of liquid nitrogen.
The reason for using a turbomolecular dry pump is to minimize the possibility
that oil and other material will backstream into the detector dewar and
contaminate the detector. This is also the reason for using a liquid nitrogen
cold trap. Actually the cold trap protects both the detector dewar from
material coming from the pump and protects the pump from contaminating
material coming from the detector dewar. Never use the pumping station
without using the cold trap.
Personal Protective Equipment:
A cold mitten is needed for handling the cold trap and liquid
nigrogen funnel. Make sure a mitten is available before beginning pumping
operations.
Pre-pumping Procedures:
-
Make sure value 3, the air inlet and the compressed gas inlets are all
closed.
-
Make sure the LN tank cold tap is clean, dried, installed, and secured
with screws with the red knurled caps.
-
Make sure valve 1 and valve 2 are open.
-
Make sure the vacuum hose attached to the high vacuum flange is secure
and properly attached to the dewar being pumped. Most of our dewars use
a standard vacuum hose attachment fixture. The dewar vacuum valve may be
closed or open, depending on whether or not the dewar has a vacuum or not.
Pumping Procedures:
-
Turn on the vacuum pump by pressing the large green button on the side
of the pump. The pump is the black metal box at the bottom of the pumping
station. The whirring sound of the pump should be heard. The pitch of the
pump sound increases as the speed of the pump increases. You should begin
to see the pressure drop, as measured by the low pressure gauge. The meter
for the low pressure gauge is mounted at the left-front edge of the pumping
station.
-
Fill the cold trap with liquid nitrogen. A funnel to aid in filling the
cold trap is kept on the pumping station table top. Use caution when handling
the funnel after the cold trap is filled since the funnel will be very
cold. Use a cold mitten or pliers to handle the funnel. Fill the cold trap
only until LN begins to bubble out the top of the trap. Overfilling the
trap may cause excess LN to spill onto some of the vacuum joints which
could freeze and result in loss of vacuum.
-
Monitor the vacuum with the low pressure gauge. The pressure typically
drops to about 10 mTorr or less on this gauge when pumping our lab dewars.
If the dewar being pumped has been open for some time you may want to pump
for 15 minutes to 1 hour to allow the dewar to outgas. Otherwise just a
few minutes of pumping at 10 mTorr should be sufficient.
Post-pumping Procedures:
-
Close the vacuum valve on the dewar.
-
Close valve 1.
-
Turn off the vacuum pump by pressing the green button on the front of the
pump.
-
Loosen, but don't yet remove the red knurled screws on the top of the cold
trap.
-
Close valve 2.
-
Open valve 3 and the air inlet valves. This brings the vacuum hose back
up to atmospheric pressure. This will be indicated on the high pressure
gauge.
-
Open valve 2. This brings the LN cold trap up to atmospheric pressure.
-
Remove the red knurled screws holding the LN cold trap cover in place and
remove the trap from the outer jacket. Caution: the trap is filled with
liquid nitrogen. Use a cold mitten when handling the trap.
-
Carefully pour the LN from the cold trap back into the LN holding flask.
-
Set the cold trap aside to warm up to room temperature. Do not put the
trap back into the outer jacket since this will cause water to accumulate
inside the pump.